A Flow Cytometric Method for Rapid Determination of Sperm Concentration and Viability in Mammalian and Avian Semen
PREBEN CHRISTENSEN*,
JENS P. STENVANG
AND
WILLIAM L. GODFREY
From the * Department of Clinical Studies, Section
for Reproduction, Royal Veterinary and Agricultural University, Frederiksberg,
Denmark;
Bartholin Instituttet, Copenhagen,
Denmark; and
BD Biosciences Immunocytometry
Systems, San Jose, California.
|
Correspondence to: Dr Preben Christensen, Department of Clinical Studies,
Section for Reproduction, Royal Veterinary and Agricultural University,
Dyrlaegevej 68, 1870 Frederiksberg C, Denmark (e-mail:
PC{at}kvl.dk). |
A new flow cytometric method has been developed to rapidly determine sperm
concentration and viability in semen from bulls and boars. Sperm viability was
determined on the basis of staining with SYBR-14 and propidium iodide (PI),
and this allowed detection of live (membrane-intact) sperm, dying (moribund)
sperm, as well as dead cells. Fluorescent microspheres (beads) were used to
determine sperm concentration. The use of SYBR-14 at 50 nM and PI at 12 µM
in combination with the FACSCount diluent in the counting tubes resulted in a
uniform staining after 2.55 minutes at room temperature. Reagent
staining was reproducible enough to allow subsequent semiautomated analysis of
data using Attractors software. In experiment 1, this method was used to
analyze semen from boars, rams, rats, rabbits, humans, and turkeys. In
experiment 2, Attractors analysis was performed by the FACSCount AF flow
cytometer, and sperm concentration determination with this system was compared
with results obtained by a spectrophotometer and an electronic cell counter,
which is routinely used by bull artificial insemination centers. When compared
to microscopic counting in a hemocytometer, the FACSCount AF flow cytometer
was two and four times more accurate than the spectrophotometer and the
electronic cell counter, respectively. In addition, the FACSCount AF flow
cytometer determined both sperm concentration (coefficient of variation 3.3%)
and sperm viability (coefficient of variation 0.7%) with high precision.
Key words: Cytometry, spermatozoa, count, vitality, precision, accuracy
This article has been cited by other articles:

|
 |

|
 |
 
S. Perticarari, G. Ricci, M. Granzotto, R. Boscolo, C. Pozzobon, S. Guarnieri, A. Sartore, and G. Presani
A new multiparameter flow cytometric method for human semen analysis
Hum. Reprod.,
February 1, 2007;
22(2):
485 - 494.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
P. Christensen, D. Boelling, K. M. Pedersen, I. R. Korsgaard, and J. Jensen
Relationship Between Sperm Viability as Determined by Flow Cytometry and Nonreturn Rate of Dairy Bulls
J Androl,
January 1, 2005;
26(1):
98 - 106.
[Abstract]
[Full Text]
[PDF]
|
 |
|
Copyright © 2004 by The American Society of Andrology.